HEART / System design

System design

Light leaves the cuff, travels through the sleeve, enters the skin, and comes back as a signal. The diagram traces that loop from left to right along the top and back along the bottom.

HEART block diagram HEART has four measurement channels. The microcontroller drives an LED driver and four LEDs, one per channel, each able to emit green, red and infrared light separately. Each LED feeds its own Corning Fibrance light-diffusing fiber through an FC connector, so the sleeve carries four Fibrance fibers that illuminate the forearm. Reflected light is collected by plastic optical fibers (POF) and returned to four photodiodes, digitised by a Texas Instruments ADC, and processed by the microcontroller, which sends readings over Bluetooth Low Energy to a phone app. A battery and power regulation block supplies the cuff. An optional removable blood pressure module is shown dashed. Cuff (electronics) Sleeve Phoneapp HR and SpO₂ Microcontroller BLE SoC (FPGA option under evaluation) LED driverswitches wavelengths 4 × LED, one per channelSFH 7018B, each emits green, red or IR FC connectorsfiber interface 4 × Fibrancelight-diffusing,one per LED Forearmtissue blood absorbslight POFreflected light 4 × photodiodeone per channel ADCTexas Instruments plastic optical fiber return BLE Battery and power regulation Optional: removable blood pressure module (pump + airbag)
Figure 1. HEART block diagram. Solid boxes are in the current design; the dashed box is under consideration. On a phone, scroll the diagram sideways.

Cuff

The cuff holds every electronic part. It switches the LEDs, focuses their light into the fibers, turns the returning light into numbers, and sends results to the wearer's device.

  • Four LEDsams OSRAM SFH 7018B, one per measurement channel. Each LED emits green, red and infrared light separately, one wavelength at a time [4]
  • Optical couplingEach LED feeds its own Fibrance fiber through an FC connector; coupling modeled in Ansys Zemax OpticStudio [5]
  • Signal acquisitionFour photodiodes, one per channel, and a Texas Instruments ADC
  • Processing and wirelessBluetooth Low Energy microcontroller; an FPGA dedicated to reading the ADC is being evaluated as an alternative

Sleeve

The sleeve is the part the wearer feels. It holds the fibers flat against the forearm and blocks outside light so the sensor only sees light that passed through tissue.

  • Emitter fibersFour Corning Fibrance light-diffusing glass fibers, one per LED, that glow along their length [3]
  • Collector fibersPlastic optical fibers (POF) that gather reflected light and route it back to the photodiodes
  • Compression fabric80–85% nylon and 15–20% elastane, blacked out, targeting about 15–20 mmHg
  • CoverageForearm only; the sleeve stops below the elbow

Interface and app

Connectors let the sleeve and cuff separate, and the app turns raw readings into numbers the wearer can understand.

  • Fiber connectorsFibrance cables with factory-fitted FC connectors, plus FC extension and LC connectorization kits from Versalume [6]
  • Wireless linkBluetooth Low Energy from the cuff to a phone or computer
  • AppDisplays heart rate and SpO2 for wellness tracking
  • FabricationCircuit boards milled in-house on LPKF ProtoMat S103 and ProtoLaser ST machines [7]

Key design decisions

Each row records what we chose, why, and what else we considered.

DecisionWhat we choseWhyAlternatives considered
Light deliveryLight-diffusing optical fibers sewn into fabricFibrance fiber is thin, bends tightly and emits light evenly along its length, so illumination can be spread across the forearm with no rigid parts on the skin.A rigid sensor patch, as in wrist wearables
Getting LED light into the fiberEach LED feeds its fiber through an FC connector, with no lens stageA lens coupling stage was modeled in Zemax and rejected because it was not cost-efficient for the light it added.Aspheric condenser lens coupling; low-cost FC-compatible lasers
Fibers per channelFour measurement channels, each with one LED and one Fibrance fiberEach LED emits green, red and infrared separately, so a single fiber carries all three wavelengths for its channel, and each wavelength is still switched and measured on its own.One fiber per wavelength (three per sleeve)
Where it is wornForearm onlyLarge, flat measurement area and easy to put on; nothing crosses the elbow joint.Full-arm sleeve; wrist band
FabricPre-made blackout compression sleeves (nylon/elastane), cut to lengthGentle compression keeps the fibers in contact with skin, and the blackout fabric keeps room light from reaching the sensor. Buying sleeves is faster and more consistent than knitting our own.Knitting or sewing a custom sleeve
ProcessorBluetooth LE microcontroller (FPGA under evaluation)We compared Nordic nRF54L15 and nRF52840 BLE SoCs; an FPGA could read the ADC faster and leave wireless to a simpler chip.FPGA plus separate radio

How the design got here

  1. Two concepts, one decision matrixWe compared a fall-detection tracker for older adults with a PPG wellness earring, scoring each on cost, market, regulatory risk and technical feasibility. The result was close: the earring led on impact, the tracker on risk.
  2. From diagnosis to wellnessReframing the optical concept from stroke detection to general wellness monitoring made validation realistic for a two-semester project and placed it under the FDA's general wellness policy.
  3. From earring to sleeveWe moved the optical sensing to a forearm sleeve with fibers in compression fabric, giving far more sensing area than an earlobe. The project became HEART.
  4. Refining the optics (current)We tried LEDs pressed directly against the fiber, considered low-cost lasers, and modeled lens coupling in Zemax, which we rejected as not cost-efficient. The design now uses four measurement channels, each with one LED and one Fibrance fiber.

Backup plans and options

If the fibers fall short

Our contingency replaces the fibers with a flexible strip of embedded LEDs and attachable sensors, keeping the same sleeve, cuff and app.

Blood pressure module